During an open thermocouple condition, some invalid values may be returned before
Note
the open thermocouple is reported.
Cold-Junction Temperature Measurement Accuracy
© National Instruments Corporation
If an open thermocouple is detected, the NI USB-TC01 software displays a
message and lists the internal CJC temperature, as shown in Figure 8.
Figure 8. NI USB-TC01 Software Open Thermocouple Message
Heat from other nearby heat sources can cause errors in thermocouple
measurements by heating up the terminals so that they are at a different
temperature than the cold-junction compensation sensor used to measure
the cold junction.
Minimizing Thermal Gradients
Thermocouple wire can be a significant source of thermal gradients if it
conducts heat or cold directly to terminal junctions. To minimize these
errors, use the following guidelines:
Use small-gauge thermocouple wire. Smaller wire transfers less heat
to or from the measuring junction.
Avoid running thermocouple wires near hot or cold objects.
If you connect any extension wires to thermocouple wires, use wires
made of the same conductive material.
9
NI USB-TC01 User Guide and Specifications