Specifications

Electromagnetic interference can adversely affect the measurement accuracy of
Caution
this product. The input terminals of this device are not protected for electromagnetic
interference. As a result, this device may experience reduced measurement accuracy or
other temporary performance degradation when connected cables are routed in an
environment with radiated or conducted radio frequency electromagnetic interference. To
limit radiated emissions and to ensure that this device functions within specifications in its
operational electromagnetic environment, take precautions when designing, selecting, and
installing measurement probes and cables.
Input Characteristics
NI USB-TC01 User Guide and Specifications
The following specifications are typical at 25 °C, unless otherwise noted.
Recommended warm-up time.................15 minutes
Number of channels................................1
ADC resolution.......................................20 bits
Input range ..............................................±73.125 mV
Common-mode range
Channel-to-USB Ground .................±30 V
Common-mode rejection ratio (0 to 60 Hz)
Common-to-USB ground ................>145 dB
Noise rejection (50/60 Hz) .....................>80 dB
Temperature measurement ranges ..........Works over temperature
Conversion time......................................250 ms
Sampling rate ..........................................4 samples per second max
Input bandwidth (–3 dB).........................1 Hz
Differential input impedance ..................20 MΩ between isolated 3.3 V
ranges defined by NIST
(J, K, R, S, T, N, E, and
B thermocouple types. The E type
has a maximum limit of 900 °C.)
and ground
10