Sterilization in boiling water
All of the components (except for the air tube and the masks) can be immersed in boiling
water for no longer than 1 minute.
Boil the components in plenty of water.
Disassemble all components of the nebulizer and boil them separately.
MAINTENANCE, STORAGE AND SERVICING
Order all the spare parts from your retailer and trusted pharmacy or contact COLPHARMA customer
service (see intro).
It is recommended to replace the nebulizer, mouthpiece, nose piece and masks after 1 year of
use.
Do not dry the nebulizer in the oven or with a hairdryer or its components may be damaged.
Always check that the filter is clean and replace it immediately if dirty or after a maximum of 60
days of use. Spare filters are provided with the device.
To replace the filter (14), remove the filter cover from its seat.
Remove the used air filter and insert the new filter. Reposition the filter cover in its seat.
Do not use the device without a filter installed. Use only original filters.
Do not wash the filter. If the filter is wet, replace it. If the filter is damp it may cause the device to
stop operating.
Wash the filter cover regularly (do not boil it) then dry it carefully before repositioning it in its seat.
Never open the compressor compartment.
Repairs must be performed exclusively by specialized personnel.