Manufacturer's declaration-electromagnetic immunity
The FocusTemp is intended for use in the electromagnetic environment (for
home healthcare and professional healthcare) specified below.
The customer or the user of the FocusTemp should assure that it is used in
the environment specified below.
Immunity
IEC 60601
test
test level
Conducted
3 Vrms:
RF
0,15 MHz –
IEC 61000-
80 MHz
4-6
6 Vrms:
in ISM and
amateur
radio
bands
between
0,15 MHz
and 80
Radiated
MHz
RF
IEC 61000-
80 % AM at
4-3
1 kHz
10 V/m
80 MHz –
2,7 GHz
80 % AM at
1 kHz
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply to all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Compliance
level
Not applicable
Portable and mobile RF
communications equipment must
Not applicable
not be used close to any parts of the
FocusTemp including cables, other
than the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
10 V/m
d = 2,3 √P 800MHz to 2,7 GHz
80 MHz – 2,7
Where P is the maximum output
GHz
power rating of the transmitter
80 % AM at 1
in watts (W) according to the
kHz
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey, a
should be less than the compliance
level in each frequency range.b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
EN-21
Electromagnetic environment-
guidance (for home healthcare
and professional healthcare
environment)