3 Operation

3.1 Liquid Handling Station

Equip the work table correctly
Make absolutely sure that the labware is placed on the work table as entered in the software/method.
Ø
In order to compensate for height differences, it is absolutely necessary to use the adapters recom-
Ø
mended by the software.
Non-observance poses the risk that the Liquid End will strike the labware and become damaged.
Ø
Only use height adapters in the designated direction. Do not apply force.
Ø
a. Switch on the device with the on/off switch on the front right of the device.
b. Start the software on the PC. A connection is established between the device and the PC.
c. During initial commissioning or after creating a new workbench, select the device type that you acquired
under System | Settings | Device Manager | Device Type.
d. Create a method or load an already existing method. This procedure is described in detail in the separate
user manual for the Liquid Handling Station operating software.
e. Open the front door and equip the work table with the labware that is required for the method and that
was selected in the software.
f. Close the front door and start the program.
Risk of crushing
Only reach into the work area when the device is at a standstill!
Monitored front door
In delivery condition, the program can only be started when the front door is closed. If the front door is
opened during a program, the operation is interrupted once the current work step has been completed.
Once the front door is closed again, the program continues from the point of interruption.
70