Caution!:
Please handel your balance with care.
Take notice of the following recommendations.
* Liquids (like water) which come in contact with the balance can harm the accuracy.
* Overload of the balance can harm the accuracy of the balance.
* Extreme temperature changes can harm the accuracy of the balance.
* A clash with a heavy machine can make the balance useless.
* Changes in the balance may only be done by your supplier.
* We recommend you to let the balance check by your supplier every year.
* Only trained employees are allowed to use the balance.
* If you warm up the balance for 15 minutes before using, the accuracy of the balance will be better.
* Always use the parts of your supplier.
Calibrate the scale
Make sure you have approved weight mass to make reliable calibration
1
Turn off the scale
2
Press and hold ZERO and turn On the scale
3
Display shows CAL and the AD vakue
4
Press UNIT the display will show flashed calibration weight
Press UNIT to confirm
5
Display shows ------ and AD value
6
Take away the weight and turn off the scale
7
Turn On the scale and check the weight
© RIBA WEEGTECHNIEK BV
16