Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 27
MP6025822-22C1 / A
• In the presence of RF interference with at least one clear channel, all operations required of
the CareLink USB and the pump functioned as intended.
Data security: The CareLink USB (MMT-7306) is designed to only accept RF communications from
recognized and linked devices. Carelink USB ensures data security via proprietary means and
ensures data integrity using error checking processes, such as cyclic redundancy checks.
Latency: The wireless function operated when commanded. There was no latency when
establishing the wireless connection or transmitting data via the wireless connection in the
adverse RF environment.

Specifications

Operating conditions
Storage conditions
Transmitter frequency
Maximum output power
(EIRP)
RF communications
Temperature: 5° C to 40° C. Humidity: 20% to 90%. Atmospheric pressure: 70.0 kPa to 106.0 kPa
Temperature: -20° C to +55° C. Humidity: 5% to 95%, non-condensing. Atmospheric pressure:
49.5 kPa to 106.0 kPa
2.4 GHz, 2M65G1D modulation
-14.0 dBm/1 MHz
Frequency range: 2.4 GHz; operating range of 1.8 meters (6 feet).
Operating Frequencies: 5 frequencies are used: 2420, 2435, 2450, 2465 and 2480 MHz
Utilizes the IEEE 802.15.4 protocol with the proprietary data format.
Bandwidth: 5 MHz which is allocated channel bandwidth per the IEEE protocol.
7

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Mmt-7306

Inhaltsverzeichnis