Herunterladen Inhalt Inhalt Diese Seite drucken

Possible Cause - EKOM DUO serie Anleitung Zur Installation, Bedienung Und Wartung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 56
DUO
To follow the rules of personal hygiene for works with contaminated material. To separate,
label, packing and providing for decontamination of contaminated parts by course of
national regulations.
To perform the repair of damaged parts.
FAILURE
Compressor does not
start
Compressor often
switches on
Prolonged
running
compressor
Compressor is noisy
(knocking, metal noises)
Suction unit does not
work, or works
intermittently
Suction unit is not
sucking, or is sucking
insufficiently, motor
works
Dryer doesn´t dry
(condensat appears in
the air)
06/2009

POSSIBLE CAUSE

No voltage in pressure switch
Disconnected winding of motor, damaged
thermal protection
Faulty capacitor
Seizure of piston or another rotary part
Pressure switch does not switch on
Air leak in pneumatic distribution system
Leaking check valve
Greater volume of condensed liquid in
pressure vessel
of
Air leak in pneumatic distribution system
Worn piston ring
Contaminated input filter
Contaminated filter in dryer
Defective solenoid valve
Damaged bearing of piston, piston rod,
motor bearing
Loose or cracked spring
No voltage in suction unit clamp
Exhauster overheating
(thermal protection switched off)
Suction line leaking, foreign object
trapped in suction piping, blocked
exhaust line
Corrupted ( long-time running at the low
pressure, age) or very polluted drying
refill
Check voltage in socket
Check fuse – replace faulty one
Loosen terminal – tighten it
Check power cord – replace faulty
one
Replace motor or re-wind it
Replace capacitor
Replace damaged parts
Check the function of pressure
switch
Check
system – seal loose joint
Clean
replace valve
Drain condensed liquid
Check
system – seal loose joint
Replace worn piston ring
Replace the contaminated filter
with a new one
Replace the output filter in a
chamber or also its refill if it gets
decomposed or if it is very dusty
Repair or change the valve
Replace damaged bearing
Replace damaged spring
Check voltage in the socket
Check fuse
Loose clamp – tighten
Check the power cord
Methodically test the wiring circuit
The check of functionality of
housing ventilator – replace the
non-functioning one
Check of the clearness of
sucking and exhausting tract -
(remedy hose cranks and
remove foreign objects)
Check/replace damaged suction
line, check/clear suction line of
foreign objects, unblock exhaust
line
Replace drying refill and filters
- 35 -
REMEDY
pneumatic
distribution
valve,
replace
seals,
pneumatic
distribution
NP-DUO-2

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

DuoDuo 2vDuo 2

Inhaltsverzeichnis