15.4.

Replacement of output filter in air dryer

Before beginning, reduce the air pressure in the tank to zero
and disconnect the appliance from the mains.
(Figr.13)
In the case of a regular operation of a dryer it is necessary to replace the dryer
filter in the upper part 1 time a year.
• Remove the dryer plug (1) by unscrewing to the left from the dryer head (4).
• Put in a new filter (2).
• Refit the dryer plug to the dryer head (4) and tighten it to the right.
15.5.

Replacement of filter in condensation and filtration unit

Before beginning, reduce the air pressure in the tank to zero
and disconnect the appliance from the mains.
(Fig.14)
In the case of a regular operation of a condensation unit it is necessary to
replace the filter inside the filter with automatic desludging 1 time a year.
• Release a safety lock (1) on the filter vessel by its pulling downwards, slightly
rotate the filter cover (2) to the left and take it out.
• Unscrew the filter holder (3) by its rotation to the left.
• Replace the filter and fix the new one by rotation of the holder to the right
back on the filter body.
• Replace the filter cover and secure it by turning to the right until the safety
pin locks.

16. STORAGE

If the compressor will not be used for a prolonged time period, drain any condensate from the air tank. Then
turn on the compressor for 10 minutes, keeping the drain valve open (1) (Fig.11). Switch off the compressor
by switch (3) at pressure switch (2) (Fig.10), close the drain valve and disconnect the appliance from the
mains.

17. DISPOSING OF THE APPLIANCE

Disconnect the appliance from the mains.
Release air pressure in the pressure tank by opening the drain valve (1) (Fig.11).
To follow the rules of personal hygiene for works with contaminated material.
To separate, label, packing and providing for decontamination of contaminated parts by course of
national regulations.
The components of the product are non-toxic.
Dispose of the appliance following all environmental regulations.
Inside parts of unit can be contaminated with biological material by reason incorrect using.
Before clearing and waste disposal pass on special institution for decontamination.

18. REPAIR SERVICE

Guaranteed and post-guarantee repairs must be done by the manufacturer, its authorized representative, or
service personnel approved by the supplier.
The manufacturer reserves the right to make changes to the appliance without notice. Any changes made
will not affect the functional properties of the appliance.

19. SOLVING PROBLEMS

Caution! Before proceeding, depressurize the air tank to zero and disconnect the appliance
from the mains.
Troubleshooting can be performed only by qualified personnel.
In case of repair of parts of device which might be contaminated please follow bellow mentioned instruction: