REPLACEMENT OF THE NEBULIZER
The nebulizer must be replaced after a long period of inactivity,
in case it shows deformations or breakings, or when the nebulizer
nozzle is obstructed by dry medicine, dust, etc. Medel recommen-
ds to replace the nebulizer after a period between 6 months and
1 year depending on the usage.
Use original nebulizers only.
!
REPLACEMENT OF THE AIR FILTER
In normal usage conditions, the air filter must be replaced appro-
ximately after 500 working hours or after each year. Medel recom-
mends to periodically check the air filter
and, if the filter shows a grey or brown colour or is wet, replace it.
Extract the filter (Pic. A-6) and replace it with a new one. Do not
try to clean the filter for reusing it.
Use original filters only.
!
Do not use the device without filter.
POSSIBLE PROBLEMS AND HOW TO SOLVE THEM
PROBLEM
SOLUTIONS
The device does not
Make sure the power plug is firmly fitted to the
switch on
wall socket.
The device does not
Make sure that the ends of the air tube are
nebulize
or
nebulizes
tightly fitted to the main unit and nebulizer.
weakly
Verify that the nebulizer is not empty or has
been filled with the proper quantity of medicine
(max 8 ml).
Verify that the nebulizer nozzle is not obstructed.
The device is noisier
Make sure the filter is assembled in the correct
way.
28
100024_MedelProfessional_USERmanual_REV.02.indd 28-29
The device is very hot
Should the device not start working properly again, address to
Medel CUSTOMER SERVICE.
MAINTENANCE AND REPAIRS
In case of failure, address to qualified personnel authorized by
Medel International Srl. Do not open the device in any case. The
unit has no user-serviceable parts within and does not need inter-
(10 – 12 treatments)
nal maintenance or lubrication.
TECHNICAL SPECIFICATIONS
Model: MEDEL Professional
Power supply: 220-230 V 50 Hz AC
Minimum filling volume: 2 ml
Maximum filling volume: 8 ml
Nebulization rate: 0,3 ml/min
Maximum pressure: 2,2 bar
Weight: 2,2 kg
Size: 210x220x240 mm
Noise level: 52 dBA
Expected service life: 2000 hours
Make sure the air-vents opening are not ob-
structed.
Class II device as regards protection against electric shocks.
Nebulizer, mouthpiece, nosepiece and masks are type BF ap-
plied parts.
Device for continuous use.Device not suitable for use in ana-