Compliance
The table illustrates the compliance of the air tubing:
Air tubing
Standard
SlimLine
ClimateLineAir
ClimateLineAir Oxy
Guidance and manufacturer's declaration electromagnetic emissions and
immunity
Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to EMC information provided in this document.
The Lumis device has been designed to meet EMC standards. However, should you suspect that
the device performance (eg, pressure or flow) is affected by other equipment, move the device
away from the possible cause of interference.
Guidance and manufacturer's declaration—electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer
or the user of the device should assure that the device is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/Flicker Emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer
or the user of the device should assure that the device is used in such an environment.
Immunity test
IEC60601-1-2 test
level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV differential
IEC 61000-4-5
mode
±2 kV common mode
26
Compliance (cm H
O/L/min) with pressure of 60 cm H
2
Air tubing with tubing elbow
1.074
0.467
-
-
Compliance
Group 1
Class B
Class A
Complies
Compliance level
±8 kV contact
±15 kV air
±2 kV
±1 kV for input/output
lines
±1 kV differential
mode
±2 kV common mode
Air tubing only
1.056
0.454
0.482
0.729
Electromagnetic environment — guidance
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage network that
supplies buildings used for domestic purposes.
Electromagnetic environment — guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.