Endress+Hauser Liquicap M Montage- Und Demontageanleitung Seite 6

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

EA0012F/00/a2/11.08
Dismounting the cable from probe
FMI51, FMI52
7
8
YE
GN/YE
9
RD
10
2
2
FTI5x
YE
GN/YE
9
RD
RD
10
Without active
build up compensation
build up compensation
Exchange of cable on Liquicap M/Solicap M - separate version
Dismounting:
7
Loosen the cable gland.
8
Unscrew the adapter disc (AF32 mm) and counter at
probe head.
9
Loosen the circlip ring with screwed wire(s) with the
circlip pliers and pull it out.
7
8
10
Dismount bunch plug with red wire, therefore
grip the nut (M4) with a pliers and pull it out.
11
For probes with active build up compensation:
Pinch off the yellow wire from the red probe cable
"Guard" below the connection part.
The installation of the new cable is described on the
following pages.
GN/YE
YE
11
RD
With active
View into the probe head
(as example: FMI51, FMI52)
Endress+Hauser Maulburg
Endress+Hauser Maulburg