Loosen the tube connector on the bottle cap.
Put a filled desiccant cartridge into the adapter on the titrant and solvent bottle caps. Put an empty
desiccant cartridge into the adapter on the waste bottle cap.
Note: The molecular sieve in the filled cartridges must be regenerated regularly depending on the humidity of the
laboratory. Refer to
Regenerate the molecular sieve
on page 19.
Push the inlet tube through the connector. Make sure that the end of the tube is at the bottom of the
bottle. Tighten the connector on the bottle cap.
Pump 1 is used to fill the measurement cell with solvent. Pump 2 is used to empty the measurement
cell. Refer to the illustrated steps that follow.
Tidy the work area
Attach the tubes to the instrument with the clips on the electrovalve and the sensor holder. Refer to
the illustrated steps that follow.
English 13