sample reference measured with a CO-oximeter. The pulse rate
accuracy has been verified by Emulator. Pulse oximeter
measurement are statistically distributed and about two-thirds of
the measurements are expected to come within the specified
accuracy range compared to CO-oximeter measurements.
SpO
range
2
SpO
Accuracy
2
(Arms)
PR range
PR accuracy
Wave length
Output power
Pedometer
Range
Storage
Capacity
Sleep analysis
Record
parameters
Record interval
Drop analysis
Mobile APP
iOS
Android
6

Electromagnetic Compatibility

The device meets the requirements of EN 60601-1-2.
Warnings and Cautions
Using accessories other than those specified in this manual may result
in increased electromagnetic emission or decreased electromagnetic
immunity of the equipment.
The device or its components should not be used adjacent to or stacked
70% to 100%
80-100%:±2%, 70-79%:±3%
30 to 250 bpm
±2 bpm or ±2%,whichever is greater
660-940nm
Red/Infrared: 3mW max. avg.
0 to 99999 steps
4 records,10 hours for each
SpO
, pulse rate, motion
2
4s
Yes
iOS 9.0 or above, iPhone 4s/iPad 3 or
above
Android 5.0 or above, with Bluetooth 4.0
BLE