Herunterladen Inhalt Inhalt Diese Seite drucken

KERR Demi Plus Bedienerhandbuch Seite 8

Vorschau ausblenden Andere Handbücher für Demi Plus:
Inhaltsverzeichnis

Werbung

The Demi
Plus
is intended for use in the electromagnetic environment specified below. The customer or user of the Demi
Guidance and manufacturer's declaration – electromagnetic emissions
EMISSIONS TEST
RD emissions
CISPRA 11
RD emissions
CISPRA 11
Harmonic emissions
Voltage fluctuations/flicker emissions
IEC 60601-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
IMMUNITY TEST
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
Voltage variations on power supply input lines
IEC 61000-4-11
Power frequency (50/60Hz) magnetic field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Note 1: At 80Mhz and 800Mhz the higher range applies
1
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radio, amateur radio, AM and FM broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. In the measured field strength in the location in which the Demi
the applicable RF compliance levels above, the Demi
Plus
Demi
.
2
Over the frequency range of 150Khz to 80Mhz, field strengths should be less than 10V/m.
Plus
The Demi
is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the Demi
mobile RF communications equipment (transmitters) and the Demi
Recommended separation distances between portable and mobile RF communications equipment and the Demi
Radiated Maximum output power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P
is the maximum output power of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80Mhz and 800Mhz the separation distance for the higher frequency range applies
8
COMPLIANCE LEVEL
Group I
Class B
Class A
Complies
COMPLIANCE LEVEL
+/-4kV contact
+/-8kV air
+/-2kV for power supply lines
+/-1kV for input output lines
+/-1kV for differential mode
+/-1kV for common mode
30% UT for 0.5 cycle
60% UT for 5 cycles
<5% UT for 5 sec
3 A/m
10Vrms
150 KHz t0 80 MHz
10V/m
80MHz to 2.5GHz
Plus
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
Plus
as recommended below according to the maximum power output of the communications equipment.
Separation distance according to frequency
m
150 KHz to 80Mhz
d = 0.4√P
0.1
0.2
0.4
1.3
4.0
Plus
shall insure it is used in such an environment.
ELECTROMAGNETIC ENVIRONMENT -GUIDANCE
The Demi
Plus
uses RF energy only for its internal function. Therefore, its RF emissions are very
low and not likely to cause any interference in nearby electronic equipment.
The Demi
Plus
is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power distribution network that supplies
buildings or for domestic purposes.
ELECTROMAGNETIC ENVIRONMENT -GUIDANCE
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Note: UT is the AC mains voltage prior to the application of the test level.
Power frequency magnetic field should be at levels characteristic of a typical commercial or
hospital environment.
Portable and mobile RF communications equipment should be used no closer to any part of
Plus
the Demi
including the power supply and cable than the recommended separation distances
calculated from the equation applicable to the frequency of the transmitter.
Field strength from fixed RF transmitters, as determined by an electromagnetic site survey,
should be less than the compliance level in each frequency range
Interference may occur in the vicinity of equipment marked with the following symbol:
Plus
can help prevent electromagnetic distance between portable and
Plus
80 MHz to 800 MHz
d = 0.4√P
0.1
0.2
0.4
1.5
4.0
1
2
.
Plus
is used exceeds
800 MHz to 2.5 GHz
d = 0.7√P
0.1
0.3
0.7
2.2
7.0

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis