WARNING
- Follow the instructions of the manual to the letter for the best measurement.
- It is advisable to stay in the same environment for at least 5 minutes before taking the
temperature.
- Do not have a bath or physical exercise before a measurement.
- While measuring avoid exposing to direct sunlight, rain or wind; do not take measurements
near source of heat or very humid places.
- Keep far from children's reach both the connection ring (4) and the disposable plastic caps (2).

BATTERY REPLACEMENT

This device is supplied with one lithium cell (CR2032 x 1). Remember to power off the unit before replacing
the battery. A malfunction may occur if the power is on when the battery is replaced. If a malfunction occurs,
restart the device. Keep the battery away from children.

PROBLEM SOLVING

Error message
ENGLISH
Problem
Device stabilization in process.
Battery is low and no more
measurements are possible.
Measurement before device
stabilization.
The device showing a rapid ambient
temperature change.
The ambient temperature is not
within the range between 10°C and
40°C (50°F ~ 104°F).
Error 5 ~ 9, the system is not
functioning properly.
Temperature taken is higer than
42.2°C (108°F).
Temperature taken is lower than
34°C (93.2°F).
Device can not be powered on to
the ready stage.
Solution
Wait until
stops flashing.
Replace the battery.
Wait until
stops flashing.
Allow the thermometer to rest in a room at
least 30 minutes at room temp. (10°C ~
40°C, 50°F ~ 104°F).
Allow the thermometer to rest in a room at
least 30 minutes at room temp. (10°C ~
40°C, 50°F ~ 104°F).
Unload the battery, wait for 1 minute and
repower it. If the message reappears,
contact the retailer for service.
Check the integrity of the probe cover and
take a new temperature measurement.
Make sure the probe cover and the lens
are clean and take a new temperature
measurement.
Change with a new battery.
6