3. When "Cal" is displayed, move the plunger slowly
in and out, until the display changes to show the
indicator measuring. This movement will calibrate
the indicator.
4. This feature will happen each time the battery
is replaced.
5. If you don't move the spindle, "Cal" will stay on the
display for only 10 seconds and the display will go
blank. To return to "Cal" mode move the spindle or
quickly press the on/off button.
6. If you move the spindle to fast it will take longer for
the tool to initialize.

NEW ZERO SEQUENCE

7. After speed mode is completed press the ZERO/
ABS button and the display will show a dashed line
as seen in Fig.6 below. The line will incrementally
disappear from left to right Fig.7; this is a visual
clue on how long you should wait for the tool to zero
out. Make sure not to move the spindle during that
time. This will happen each time the tool is zeroed,
takes about a second to complete.
8. Fig.7 shows a representation of how the display will
look when the sequence is completed.
Fig. 6