MAINTENANCE
Exchanging the heater.
Use a wrench to unscrew the cover.
Move back the spring and the cover. Pull the
connector from the socket to disconnect the heater
from the station.
O-ring ø17 x ø1,8
Follow this process inverted to re-connect the heater.
Exchanging the resistor from the heater.
1. Move back the spring. Untighten the screws.
2. To take out the resistor, press down the lower
part of the heater on a fixed surface.
3. Separate the resistor from the heater's cable.
4. Connect the new resistor, pushing it's extreme.
5. Fix the screws tightly to avoid air-loss which
could reduce the resistor's lifetime. Finally put
the spring back in it's place.