Service Life

Service life of the suction unit is 5 years.
Service life of the battery is 200 discharge cycles.
Service life of the reusable container is 30 autoclave cycles at 121°C.
Service life of the disposable container is 30 dishwasher cycles.
TROUBLESHOOTING
NOTE– Your DeVilbiss Suction Unit contains no user-serviceable parts. If you believe your unit is not working properly, BEFORE YOU RETURN IT TO THE HOME MEDICAL
EQUIPMENT PROVIDER WHERE YOU PURCHASED IT, please take a few moments to check for these possible causes:
PROBLEM
Unit does not turn on, but green external power light is
illuminated.
Pump runs, but there is no vacuum.
Low vacuum.
Battery will not charge but charge light is illuminated.
PROVIDER'S NOTES
No routine calibration or service is required provided the device is used in accordance with the manufacturer's directions. In case of a change of patient, the device must be
reconditioned to protect the user. Reconditioning must only be carried out by the manufacturer or service provider. Between patients:
1.
Visually inspect unit for any damage, missing parts etc.
2.
Ensure that unit and accessories are clean.
3.
Use an independent vacuum gauge to verify the unit provides the proper vacuum level as stated in Specifications.
4.
Discard and replace collection container, filter, tubing, and carrying case between patients.
5.
Wipe the surface using a clean cloth dampened with disinfectant.
SE-7305
ACTION
1. Check power sources and connections.
2. Ensure wall outlet is live by plugging in a lamp.
3. If running from an internal battery, ensure that your unit has a battery installed.
4. If battery is installed, check that it is fully charged.
1. Check that all tubing is connected properly.
2. Check tubing connections for breaks or leaks.
3. Ensure that float shut-off is not activated.
4. Check for leaks or cracks in container assembly.
1. Use vacuum adjustment knob to increase vacuum level.
2. Check system for leaks.
3. Push vacuum adjustment knob in toward unit and then release.
1. Ensure that unit is equipped with an internal battery by contacting your DeVilbiss equipment provider.
2. Verify that charge light turns on.
3. Check electrical connections during charging.
4. Ensure wall outlet is live by plugging in a lamp.
EN - 9