Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The CEFAR PERISTIM PRO Electrotherapy System is intended for use in the electromagnetic environment specified below. The
customer or the user of the CEFAR PERISTIM PRO Electrotherapy System should assure that it is used in such an environment
Immunity Test
Electrostatic discharge
±6kV contact
(ESD) IEC 61000-4-2
±8kV air
Electrical fast transient/
±2kV for power supply lines
burst IEC 61000-4-4
±1kV for input/output lines
+ 1kV differential mode (line
Surge
to line)
IEC 61000-4-5
+ 2kV common mode (line
to ground)
<5% U T (>95% dip in U T ) for
Voltage dips, short
0,5 cycle
interruptions and voltage
40% U T (60% dip in U T ) for
variations on power supply
5 cycles
input lines
70% U T (30% dip in U T ) for
25 cycles
IEC 61000-4-11
<5% U T (>95% dip in U T ) for
5 sec
Power frequency
3 A/m
(50/60Hz) magnetic field
IEC 61000-4-8
NOTE: U T is the AC mains voltage prior to application of the test level.
IEC 60601
Compliance Level
Test Level
±6kV contact
±8kV air
Not Applicable - Battery
powered
Not Applicable - signal lines
less then 3 meters
Not Applicable - Battery
powered
Not Applicable - Battery
powered
3 A/m
Electromagnetic Environment -
Guidance
Risk assessment on the CEFAR
PERISTIM PRO Electrotherapy
System indicates the compliance
levels claimed are acceptable
when ESD-precautionary measures
are taken.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the CEFAR PERISTIM PRO
Electrotherapy System requires
continued
operation during power mains
interruptions, it is recommended
that the CEFAR PERISTIM PRO
Electrotherapy System be
powered from an uninterrupted
power supply.
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
27