Calibration And Adjustment - DeFelsko PosiTest PC Powder Checker Kurzanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

1. Zero the probe (pg. 4).
2. Spray powder onto a flat, smooth metal sample.
3. Measure the uncured powder with the PosiTest PC.
4. Cure the powder.
5. Measure cured thickness with a magnetic or eddy current
coating thickness gage such as the PosiTest DFT or
PosiTector 6000.
6. Compare results.
A successful verification requires that the average of a series of
measurements by both instruments be within their combined
tolerances. If not, adjust the PosiTest PC using one of the
methods described in Adjustment Techniques (pg 8).

Calibration and Adjustment

Verification of Accuracy
7