Troubleshooting
Operational
Message
Description
Unknown Error
Calibration not able to
Occurred
complete for an unknown
reason
Calibration
The M200 does not see
Unit Not
the calibration unit as
Connected
connected
Welding
The welding cable is not
Cables Not
connected to the M200
Connected
and the calibration unit
Electrode
The Electrode cable (red)
Cable Plugged
is connected to the wrong
into Wrong
range connector (low,
Connector
medium or high) on the
calibration unit
Input/Output
The wire harness (Input/
Wire Harness
Output connector) is
Not Connected
connected to the M200
but not to the calibration
unit
Exceeded
The M200 is making
Maximum
adjustments outside
Offset
of normal range There
may be problems with
calibrating the M200
The current calibration for
the level last attempted
Calibration Not
was not successful There
Completed
may be problems with
calibrating the M200
8
Remedy
Repeat the
calibration process
If the error
repeats, contact
your authorized
Swagelok
representative for
assistance
Verify the wire
harness and
welding cables are
connected
If connected, go to
Setup/Operation tab
on the M200 and
ensure "Calibration
Unit" is selected
as the connected
accessory
Connect the
welding cable to
the M200 and the
calibration unit
Connect the
Electrode cable to
the range connector
specified on the
M200 screen
Connect the wire
harness to the
calibration unit
Repeat the
calibration process
If the error
repeats, contact
your authorized
Swagelok
representative for
assistance
Repeat the
calibration process
up to two more
times If the error
repeats, contact
your authorized
Swagelok
representative for
assistance
Fig. 14 — Error Message