Direct Connection

Connect the transducer directly to the Milltronics transceiver using the two conductor shielded
cable.
When connecting to a SITRANS LUC500, a MultiRanger 100/200, or a HydroRanger
Note:
200, the white, black, and shield wires are all connected separately. DO NOT tie the white
and shield wires together.
Coaxial Extension
Connect the transducer to the Siemens Milltronics transceiver via a junction box and RG–62 A/
U coaxial cable. This setup is effective for combined runs up to 365 m (1200 ft).
Extend cable using RG-62 A/U coax.
When connecting to a SITRANS LUC500, MultiRanger 100/200, and HydroRanger
Note:
200, do NOT use coaxial cable; see diagram below for proper procedure.
7ML19985QK82
ST-H Transducer – OPERATION MANUAL
black
shield
white
white
shield
Page EN-11