Herunterladen Inhalt Inhalt Diese Seite drucken

Specifications - Philips Respironics Wisp SE Nasal Mask Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Attaching the Cushion
1. Push the cushion into the frame towards the front and press into place l.
2. The bottom of the cushion will match the alignment cutouts of mask frame m.
Attaching the Standard Elbow and Tubing
Push the standard elbow C into the cushion until the quick release tabs click into place k.
Attaching the Headgear
1. Hold the crown strap upward.
2. Make sure the wide strap marked with "Philips Respironics" is outward.
3. Slide the headgear tabs into the longer frame arm slots.
4. Connect the headgear clips to the frame or slide the tabs into the shorter frame arm slots if not
using the headgear clips.
System One Resistance Control
This symbol
may appear on your Philips Respironics therapy device or mask. The System One
setting is not applicable to this mask, or any masks requiring a separate/additional exhalation device.

Specifications

Warning: The technical specifications of the mask are provided for your healthcare professional
to determine if it is compatible with your CPAP or bi-level therapy device. If used outside these
specifications, or if used with incompatible devices, the mask may be uncomfortable, the seal of
the mask may not be effective, optimum therapy may not be achieved, and leak, or variation in
the rate of leak, may affect device function.
Pressure Flow Curve
70
60
50
40
30
20
10
0
0
5
Resistance
Drop in Pressure at
50 SLPM
All sizes
0.9 cm H
Deadspace Volume
P
104.6 mL
S/M
115.9 mL
L
140.0 mL
XL
151.3 mL
Sound Levels
A-weighted Sound Power Level 17 dBA
A-weighted Sound Pressure Level at 1 m 9 dBA
Disposal
Dispose of in accordance with local regulations.
10
15
Pressure cm H
2
100 SLPM
O
3.6 cm H
2
2
4
20
25
O
O
30

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis