Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

10.2
Disinfection
The smartbox is constructively only suitable to be disinfected by wiping.
The unit is not allowed to be inserted in liquids. The camera head is
watertight. The plug of the camera head is not allowed to encounter
liquids.
Generally, we recommend using sterile drapes for single use for the steri-
le covering of camera head and camera cable.
10.2.1 Disinfecting the camera head with liquids
Insert the camera head and cable without plug in a disinfectant
(connector may not enter disinfectant). Please follow the
recommendations of the manufacturer of the disinfectant regarding
concentration of the liquid and disinfecting time. Inserting the
camera head for more than the recommended time will not impro-
ve the result of the disinfection.
After disinfecting, rinse the camera head and cable, then wipe them with
a cloth and dry the exterior surfaces carefully.
10.3
Sterilization
The smartbox is not allowed to be sterilized. The camera head is not
suitable for sterilization with ETO.
Steam autoclavation destroys the camera head.
11

Maintenance

The camera does not have to be maintained regularly.