Connecting the light guide
• Attach the fiberoptic cable to the endoscope and turn-on the light source
with light intensity at low.
Focusing the image
• To focus the image, turn the optical focusing ring (12) on the camera head.
Minimizing the Moiree effect using fiberscopes
• Applying the Anti-Moiree version of the Compact Camera first adjust
focus ring (12) without fiberscope to get a sharpe picture of an object in
a distance of about 3 m. Then connect fiberscope with focus ring, focus
the fiberscope and turn fiberscope relatively to focus ring in order to opti-
mize the picture.
White balance
• In order to white balance the camera system, aim the endoscope at a
white object, such as a sterile gauze.
• Press combination button (13) at camera head for a short time (approx.
1- 2 sec.) and hold the endoscope focused to the white object. The image
will change to an accurate white. This procedure will allow the camera to
reproduce all colors naturally.
29