ENGLISH
2
Contraindications
All air mattress systems are contraindicated for patients
with cervical traction or unstable:
Spinal fractures
Spinal cord injury
Also:
Fractures at risk of complication by a moving
support surface.
Trauma patients where spinal injuries have not
been excluded or 'cleared'.
CAUTION
For patients undergoing cervical traction the use of the LAL
mode is contraindicated and the use of LAL and PT modes
should only be under the direction of a suitably qualified
physician.
Before placing a patient on an air mattress, a risk assess-
ment by a suitably qualified person should be completed,
to ensure the appropriateness of the support provided.
3
Technical Specifications
Electrical Specifications
NOTE
The control unit Power Inlet is used as the means of
isolating the equipment from the supply mains on all poles
simultaneously.
Parameter
Input Voltage AC
Input Frequency
Maximum Power Consumption
Circuit Protection
Fuse Type
Breaking Capacity
Mode of Operation
ProphyCair & ProphyCair+
Version: 01
Value
90 ~ 240 VAC
60 / 50 Hz
180 W ± 30 W
Dual fused, 250V, 5A
Fast blow fuse(s), std.
fuses
Bussmann S500-5-R
@125 VAC is 10kA
@250 VAC is 200A
Continuous
Performance Specifications
Parameter
Weight capacity
Standard replacement mattre-
ss
Pressure zones
Max. Flow
Support Surface Inflation Time
Patient Comfort Pressures / Alternating Pressure
Parameter
Soft Pressure
Firm Pressure
Patient Contact
Control unit and the mattress are lead free, mercury free
and latex free components.
Dartex top sheet is Halogen-free (bromide-free).
Mechanical Specifications
Parameter
Dimensions LxWxH
ProphyCair
ProphyCair +
ProphyCair Lo
ProphyCair Lo +
Weight
Power cord
Connection
Packaging
PLEASE CLEAN FILTER EVERY 3 MONTHS OR
WHENEVER DIRTY
Remove 2 thumb screws and separate filter foam from
filter cover. Wash filter foam using soap and water. Dry and
replace filter and filter cover on the unit and
fasten screws.
Support Surface
Air cushions are high frequency heat sealed, liquid proof
and washable. Base is liquid proof and washable.
Top Sheet is low friction, low shear, low force producing,
breathable, liquid resistant and highly vapor permeable.
Material is 70 Denier nylon or Dartex and a bottom foam
or air pad on the bottom of the air cushions enclosed in a
cover.
Value
455 kg
5
1275 LPM / 45 CFM
Within 120 seconds
(Max Inflate)
Value
7 ± 5 mmHg
32 ± 5 mmHg
Value
200 x 87 x 20,5 cm
200 x 99 x 25 cm
200 x 87 x 18 cm
200 x 99 x 18 cm
4 kg (CU), 7 kg
(mattress)
600 cm
1.27 cm flow single
quick disconnect co-
nnector
1 piece per box
5/16