• Operation and use

Once installation has been completed, the unit can be
switched on by pressing the main switch 1 located on the
front panel; the switch will light up and aspiration will start
by lifting one of the terminals from its seat. If you open
the cabinet door the electrical circuit is open and the
aspirator stops. In standard assembled units aspirated
air is ex haust ed through the silencer (2) (Fig. A). In order to
convey exhausted air outside you need to fi t an ex ten sion
to hose 3C and drive it outside (Fig. A).
Most of the noise and bacteria will be carried outside toge-
ther with air; we can also supply a certifi ed bac te ri al fi lter
for exhausted air.
When installing Aspi-Jet 8
directions and regulations men tioned, the engineers shall:
- connect the water supply, without removing the antispray
tube 15 (Fig. B), which protects the Rilsan tube against
bursting;
- check any possible leaking, especially near parts subject
to tension;
- adjust water pressure to a max of 4 bar by using the pressure
control device 12 (Fig. B).
• Operation and use
Aspirated liquids and air are conveyed to the manifold (4)
through the tip and lifted hose and into the canister through
tube 3A (page 15); inside the canister liquids are separated
from air.
The air is driven all the way to the motor (through pipe 3B)
and is then ex haust ed, while liquids are heavier and are
collected at the bottom of the canister.
Aspi-Jet 6
is equipped with a canister (5) whose capacity
allows 8/10-hour continuous service before reaching
max i mum level, therefore the canister needs emptying every
evening after work. Aspi-Jet 7
valve (6) located at the bottom of the canister is kept close
by depression when suction is performed; on the contrary,
when all terminals are on their seats suction stops and liquids
are drained out of the canister. In case the canister should
fi ll up during surgical operations, the probes (7) will sense
maximum level at about 3/4 of the canister (shortest-probe
level); the elec tri cal circuit opens and suction stops, while
a yellow-light indicator located on the front panel warns
that the canister is full. For Aspi-Jet 6
the unit off and empty the canister manually; for Aspi-Jet
7
, as mentioned above, the draining valve opens and the
draining pump (8) (page 15) starts working. In a few seconds
the canister is empty and suction starts again automatically.
and 9
, besides all general
has automatic drainage: a
you need to switch
12
15
Fig. B
Disinfecting antifoaming agent for
dental aspirators
Fig. B1
Puli-Jet plus new with anti-scale
agent
Fig. B2
Fig. B3
19