Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to
maintain a device in accordance with the instructions may invalidate the device's warranty. Furthermore,
failure to maintain a device may compromise the clinical condition or safety of users and/or their
caregivers. Do not perform service and maintenance while the user is in the product.
1st Day
• Read the User Guide carefully before
use and store it in a safe place for future
reference
• Store the enclosed tools on/at the product
Daily
• Use a dry cloth to clean the product
• Visual inspection. Repair or replace any
damaged or worn-out parts
• Ensure that all fasteners are in place and
fitted correctly
• Check all fittings and buckles for signs of
worn out parts
• Check that all cables connecting the power
system are in good shape
Weekly
• Castors are moving freely and all wheel
locks/brakes are easily activated
• Wipe off any residue and dirt from the
product using a cloth with warm water and a
mild detergent/soap and dry before use
• Disinfect the product. please check the
disinfection section in this User Guide for
more information
Monthly
• All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
• Oil the swing-away parts. We recommend
you use a professional lubricant system
Annually
• Inspect the frame for cracks or signs for
worn-out parts and carry out annual service.
Never use a product which appears to be
substandard or faulty
Washing
The cushion covers can be washed at 60°C
using a mild detergent in a washing machine.
The foam inserts can be washed at 40°C using a
mild detergent in a washing machine.
During and after washing the foam, do not bend,
squeeze or wring it as it may negatively affect the
shape
Warewashing machine
This product can be washed at:
• 60°C in 15–20 minutes
.. using a mild detergent in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product.
support.R82.org
Before washing
Avoid degreasing of the piston rods, the
actuators should be retracted to minimum stroke
and without load before washing.
Remove the following items and clean them
separately:
• Accessories
• Cushions
• Power unit including; Actuator, battery and
handset
After washing:
• Replace the power column going backwards
through the procedure give on page 25
• Check that the handset functionalities works
correct
Disinfection
The product without accessories and cushion
can be disinfected with a 70% disinfectant IPA
solution. It is recommended to wipe off any
residue and dirt from the product, using a cloth
with warm water and a mild detergent/soap
without chlorine and let it dry before disinfection.
Service interval
A detailed inspection of the product must be
made every 12 months (every 6 months for heavy
users) and every time the product is reissued for
use. The inspection should be carried out by a
technically competent person who has been
trained in the use of the product.
The lifetime of this product in normal use is 5
years if all maintenance and servicing is carried
out in accordance with the manufacturer's
instructions and demonstrably recorded.
Service Check List:
Check and adjust the following functions as
per their intended use. Repair or replace any
damaged or worn-out parts. Please check the:
• Seat plane adjustment. The seat plane
is adjustable and stays in position after
adjustment.
• Frame is adjustable in height
• Adjustment options for the foot rest, it is
possible to lock the angle and fasten the
telescopic tubes in position
• Seat back is adjustable and can be locked
in position
• Nuts and bolts on the product are tightened
• Oil the swing-away parts
• Product for cracks or signs for worn-out
parts
• All labels on the product are intact