Herunterladen Inhalt Inhalt Diese Seite drucken

Dentsply Maillefer X-SMART PLUS Gebrauchsanleitung Seite 47

Endo-motor
Vorschau ausblenden Andere Handbücher für X-SMART PLUS:
Inhaltsverzeichnis

Werbung

Cables and
Maximum
accessories
length
Motor Handpiece
1.65 m
cord
AC Adapter
1.80 m
Recommended separation distances between portable and mobile RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output power
of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
FISDR-JAP / F1902124.X / 06 / 2011 – updated 10/2015
 
Shield
Connector
Unshielded
Plastic
Unshielded
Plastic
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1,2 √P
0.12
0.38
1.2
3.8
12
Complies with
RF emissions, CISPR11, Class B/ Group 1
Harmonic emissions, IEC 61000-3-2, Class A
Voltage fluctuations/ flicker emission
IEC 61000-3-3
Electrostatic discharge (ESD) IEC 61000-4-2
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage
variations on power supply input lines
IEC 61000-4-11
Power frequency (50/60Hz) magnetic field
IEC 61000-4-8
Conducted RF IEC 61000-4-6
Radiated RF
IEC 61000-4-3
(m)
80 MHz to 800 MHz
d = 1,2 √P
0.12
0.38
1.2
3.8
12
800 MHz to 2,5 GHz
d = 2,3 √P
0.23
0.73
2.3
7.3
23
- 47 / 48 -

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis