The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device
should ensure that it is used in such an environment.
Immunity test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/output
Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5
± 2 kV line(s) to earth
< 5%
for 0,5 cycle
Voltage dips, short
40%
interruptions and
for 5 cycles
voltage variations
on power supply
70%
input lines
for 25 cycles
IEC 61000-4-11
< 5%
for 5 sec
Power frequency
(50/60Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE:
U
is the a.c. mains voltage prior to application of the test level.
T
FISDR-JAP / F1902124.X / 06 / 2011 – updated 10/2015
 
Guidance and manufacturer's declaration - electromagnetic immunity
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/output
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
U
(> 95% dip in
U
)
< 5%
T
T
for 0,5 cycle
U
(60% dip in
U
)
40%
T
T
for 5 cycles
U
(30% dip in
U
)
70%
T
T
for 25 cycles
U
(> 95% dip in
U
)
< 5%
T
T
for 5 sec
3 A/m
Compliance level
Floors should be wood, concrete or
ceramic tile.
If floors are covered with synthetic
material, the relative humidity should
be at least 30%.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
U
(> 95% dip in
U
)
T
T
Mains power quality should be that of
a typical commercial or hospital
environment.
U
(60% dip in
U
)
T
T
If the user of the device requires
continued operation during power
U
(30% dip in
U
)
mains interruptions, it is recommended
T
T
that the device be powered from an
uninterruptible power supply or a
U
(> 95% dip in
U
)
battery.
T
T
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
Electromagnetic environment -
guidance
- 45 / 48 -