Guidance and manufacturer's declaration - electromagnetic immunity
ABPM-100 is intended for use in the electromagnetic environment specified below.
The customer or the user of ABPM-100 should assure that it is used in such an environment.
Immunity test
IEC 60601test level Compliance level
Radiated RF
3 V/m 80 MHz
IEC 61000-4-3
to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which ABPM-100 is used exceeds
the applicable RF compliance level above, ABPM-100 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating ABPM-100.
b: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and ABPM-100
ABPM-100 is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of ABPM-100 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and ABPM-100
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
150 kHz to 80 MHz
d = 1.2√P
0,01
0.12
0,1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and reflection from structures, objects and people.
12
/ 74
ABPM-100 EN Instruction manual
Electromagnetic environment - guidance
3 V/m
Portable and mobile RF communications equipment should
be used no closer to any part of ABPM-100, including cables,
than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 2.3√P 800 MHz to 2,5 GHz
Where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and
d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey
the compliance level in each frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
should be less than
a
b
m
800 MHz to 2,5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23