3.0 Mounting the Docking Station on the VHM Wall Mount
3.1 Pull the plunger at the front of the mounting plate. Slide the Mindray V Dock mounting adapter into the mounting plate
until the plunger snaps into the clearance hole in the adapter. Tighten the (4) Nylon Screws on the underside of the plate.
Plunger
Mounting Adapter
Mounting Plate
4.0 Operating and Adjusting the VHM Wall Mount
Check pivot, tilt, and front swivel mechanisms for proper tension. Refer to the applicable section and follow the
adjustment procedure.
4.1 Adjusting Counterbalance
WARNING: Do not attempt counterbalance adjustment unless the instrument and accessories are
mounted on the Arm. Use caution while performing this procedure.
The VHM Arm must be adjusted to counterbalance the mounted instrument throughout the Arm's vertical range of motion.
When properly counterbalanced, the VHM Arm will maintain its height without engagement of the Height Locking Knob.
Because instrument weights vary, adjustment will typically be required to achieve optimal performance. A 1/2" [13mm]
socket wrench is provided for this critical adjustment.
To adjust counterbalance:
4.1.1. Grasp the Arm behind the mounted instrument(s) and carefully move the Arm to a horizontal position that allows
access to the Adjuster Cover (below left). Lock Arm in horizontal position.
4.1.2. Open the Adjuster Cover by inserting a flat blade screwdriver in the slot at the rear of the Cover and prying upward.
If necessary unlock and readjust the Arm until the Counterbalance Adjuster bolt is accessible through the Adjuster Cover
(below center). Lock Arm when Counterbalance Adjuster is accessible.
DU-MND–0013-01 Rev A 9/28/10 GCX Corp. Page 4 of 10
Nylon Screws (4)