Herunterladen Inhalt Inhalt Diese Seite drucken

NSK Multi Pad Bedienungsanleitung Seite 14

Multi pad for nlx/imd
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 20
9–1–3 [3. Lamp] : For setting the function of the LED.
At setup mode, select "Lamp" on the screen. Then you will see the next selection shown below diagram.
3.5V(NLX) / 3.3V(iMD).
sec.
1.Language
2.Gear ratio
3.Lamp
On / Off
4.Endo mode
Intensity
5.Air
Delay timer
6.Endo alert
7.Others
8.Version
NOTICE
The setting items of "Intensity" and "Delay timer" will be displayed only when lamp setting is
"On".
9–1–4 [4. Endo mode] : Endo (endodontics) mode setting.
This setting is not applicable to the NLX nano motor.
1.Language
2.Gear ratio
3.Lamp
4.Endo mode
5.Air
Display setting
6.Endo alert
7.Others
Auto REV time
8.Version
Auto FWD time
9–1–5 [5. Air] : Calibration for air driven foot control.
Air pressure may vary on the dental chair unit. The foot control of the Multi Pad system has a feature for
synchronizing the motor speed with the air pressure.
Adjust the air pressure within a range shown below by adjusting foot control. When a message of "under" is
displayed, air pressure is too low. When a message of "over" is displayed, air pressure is too high.
(4.3 – 14.5 psi or 0.30 – 1.00 bar, 0.030 – 0.100MPa).
(29.0 – 58.0 psi or 2.00 – 4.00 bar, 0.200 – 0.400MPa).
1.Language
2.Gear ratio
3.Lamp
4.Endo mode
5.Air
Display setting
6.Endo alert
7.Others
Motor MIN pressure
8.Version
Motor MAX pressure
On / Off
Case of iMD
Ncm
Nmm
%
1.0 – 3.0 sec
psi
bar
MPa
13

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis