Baumer Z-PA-EI-H Betriebsanleitung Seite 23

Programmierzubehör für inkrementale drehgeber
Inhaltsverzeichnis

Werbung

1.6) Output Levels
GB
1.7) Reference Type
1.8) Pulse Sequence
1.9) Firmware Version
2.)
Referencing (Set Reference)
2.1) Set Ref. Pulse to Shaft Pos.
3.)
Write pulse number (Write Pulses)
3.1) Write Pulses
Baumer_Handheld-Programming-Tool-Z-PA-EI-H_OI_DE-EN_1708.pdf
Delivered is the current signal output level of the connected
encoder.
Delivered is current width and blanking of the reference
pulse of the connected encoder.
Delivered is the current pulse sequence A leading to B
at clockwise shaft rotation (when looking at the encoder
flange).
Delivered is the current 6-digit firmware version.
Important: The reference pulse is set to an encoder-specific
position once having executed the functions described un-
der 3.1; 5.1; 6.1. Step as under 2.1 Set Ref. Pulse to Shaft
Pos. is to be executed last.
Only execute the functions whilst encoder shaft standstill.
Function as under 1.2 Read Position is to deliver 0 ± 1
after the programming operation.
Delivered is the currently defined ppr number of the connec-
ted encoder.
Changing the ppr number is by the arrow keys and entry
confirmation by Enter. The maximum ppr number depends
on the encoder type (see product data sheet).
Note: Encoder types ITD2P and HS35P differentiate bet-
ween channel A and B.
Explanations on functionality
 23