Difference between free-field sensitivity level G
on ear simulator with adapter (plate) and conical ring according to IEC 60318-1:2010
Mean value on B&K 4153
Frequency in Hz
in dB
100**
-12.6
125
-12.2
160*
-11.7
200*
-11.3
250
-10.9
315*
-10.9
400
-11.0
500
-7.8
630*
-5.4
800*
-3.0
1000
-0.8
1250
0.3
1600
-0.7
2000
-2.1
2500
-3.0
3150
-5.2
4000
-5.4
5000
-11.7
6300
-7.5
8000
-10.1
Difference between free-field sensitivity level G
on 6 ccm coupler according to IEC 60318-3:2010
Mean value on B&K 4152
Frequency in Hz
in dB
100**
-5.8
125
-5.8
160*
-5.7
200*
-5.6
250
-5.6
315*
-6.3
400
-7.1
500
-5.7
630*
-7.2
800*
-8.7
1000
-10.2
1250
-11.3
1600
-13.4
2000
-14.8
2500
-15.2
3150
-14.6
4000
-14.0
5000
-13.5
6300
-9.2
8000
-3.3
*
interpolated values
** extrapolated value
*** rounded to nearest 0.5 dB
and ear simulator sensitivity level G
F
Standard
Median value***
deviation in dB
on B&K 4153 in dB
-12.0
4.4
-12.0
-11.5
-11.5
3.9
-11.5
-11.0
4.3
-10.0
3.5
-7.5
-5.0
-3.0
3.4
-1.0
4.7
0.0
3.2
-0.5
4.0
-2.0
3.0
-3.0
3.1
-6.0
3.3
-4.5
3.3
-10.5
3.9
-7.0
4.5
-10.0
and coupler sensitivity level G
F
Standard
Median value***
deviation in dB
on B&K 4152 in dB
-5.0
4.4
-5.5
-5.5
-6.0
3.9
-6.0
-6.0
4.3
-6.0
3.5
-5.0
-7.0
-8.5
3.4
-10.5
4.7
-11.5
3.2
-13.5
4.0
-15.0
3.0
-15.0
3.1
-15.5
3.3
-13.0
3.3
-12.0
3.9
-8.5
4.5
-3.0
C
Interquartile
range in dB
5.5
4.5
7.0
3.5
3.5
6.0
5.5
7.0
3.0
4.0
6.0
2.5
6.0
7.0
C
Interquartile
range in dB
5.5
5.0
7.0
3.5
4.0
6.0
5.0
5.5
3.0
4.0
6.0
2.0
6.0
7.0

Quicklinks ausblenden: