Herunterladen Inhalt Inhalt Diese Seite drucken

Sennheiser HDA 300 Bedienungsanleitung Seite 2

Audiometric headphone
Inhaltsverzeichnis

Werbung

Sensitivity level
on ear simulator with adapter (plate) and
conical ring according to
IEC 60318-1:2010*
Mean value
Frequency in Hz
on B&K 4153
in dB re 1 Pa/V
63
34.9
80
35.6
100
36.1
125
36.1
160
37.2
200
38.0
250
38.0
315
38.6
400
38.2
500
37.2
630
35.4
750
33.5
800
32.7
1000
30.1
1250
27.5
1500
25.2
1600
24.3
2000
21.1
2500
20.3
3000
20.0
3150
19.8
4000
16.3
5000
26.1
6000
26.4
6300
26.7
8000
28.1
10000
19.4
12000
24.1
12500
20.2
14000
17.5
16000
16.1
* measured with additional adapter for a quicker and more accurate positioning on coupler plate
This is optionally available from Sennheiser.
Measuring conditions
Temperature:
Relative humidity:
Standards
Headphones comply with IEC 60645-1, IEC 60645-2, ANSI S3.6-2010
Standard
deviation in dB
0.58
0.75
0.58
0.88
0.40
0.31
0.67
0.77
1.03
1.05
0.75
0.61
0.53
0.28
0.37
0.26
0.33
0.52
0.54
0.55
0.55
0.46
0.20
0.25
0.36
1.56
0.43
0.81
0.68
1.04
0.85
22 °C
50 %
on 6 ccm coupler according to
IEC 60318-3:2010
Mean value
Standard
on B&K 4152
deviation
in dB re 1 Pa/V
24.5
1.08
27.2
0.90
29.1
0.68
30.5
0.52
31.7
0.41
32.5
0.29
33.1
0.44
34.2
0.27
35.3
0.22
36.3
0.22
37.5
0.28
38.2
0.41
38.4
0.34
38.8
0.61
37.4
0.62
36.0
0.29
35.3
0.20
33.2
0.58
32.0
0.42
29.7
0.36
29.0
0.29
26.7
0.14
27.0
0.36
28.9
0.54
28.3
0.54
18.4
0.74

Quicklinks ausblenden: