Herunterladen Inhalt Inhalt Diese Seite drucken

Sennheiser HDA 300 Bedienungsanleitung Seite 3

Audiometric headphone
Inhaltsverzeichnis

Werbung

Reference equivalent threshold sound pressure level
on ear simulator with adapter (plate) and conical ring according to IEC 60318-1:2010
Mean value
Frequency in Hz
on B&K 4153
in dB re 20 µPa
125
26.2
250
20.1
500
8.6
750*
5.1
1000
2.7
1500
3.2
2000
0.5
3000
-1.6
4000
0.1
5000
11.3
6000
20.9
8000
23.1
9000
27.1
10000
18.5
11200
22.9
12500
27.0
14000
32.8
16000
47.7
Reference equivalent threshold sound pressure level
on 6 ccm coupler according to IEC 60318-3:2010
Frequency in Hz
125
250
500
750*
1000
1500
2000
3000
4000
5000
6000
8000
*
interpolated values
*** rounded to nearest 0.5 dB
Standard
deviation in dB
3.9
3.1
4.0
4.6
7.1
4.5
6.0
5.4
4.4
7.2
6.1
5.4
5.7
5.6
5.5
6.9
14.4
Mean value
Standard
on B&K 4152
deviation in dB
in dB re 20 µPa
22.3
3.9
14.7
3.1
7.4
4.0
9.5
11.0
4.6
14.8
7.1
12.3
4.5
8.0
6.0
10.0
5.4
12.3
4.4
23.0
7.2
15.9
6.1
Median value ***
Interquartile
on B&K 4153 in dB
range in dB
27.0
5.5
20.0
4.0
8.0
3.5
4.5
2.0
4.5
3.0
6.5
0.0
7.0
-3.0
7.5
-0.5
5.5
10.5
6.0
21.0
7.5
23.0
6.0
27.5
7.0
18.0
6.0
22.0
8.5
27.0
10.0
33.5
7.5
45.5
12.5
Median value ***
on B&K 4152 in dB
23.0
14.5
6.5
9.0
10.5
14.5
12.0
6.5
9.5
11.5
23.0
7.5
16.0
6.0
Interquartile
range in dB
5.4
4.0
3.5
4.0
7.0
7.0
7.5
5.5
6.0

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis