27.0 Inspecting and Reissuing of Equipment

Most Sunrise Medical Seating Solutions Ltd. products are individually assessed for a client prior to issue. They are handed over in a
controlled way to ensure optimum fit and specification, checks are completed and any additional individual verbal instructions given.
We therefore recommend the following points are adhered to prior to any re-issue of equipment:

Cleaning

Follow a rigorous process of cleaning in order to decontaminate the product and eliminate the chance of cross infection.

Assessment of Fabrics / Foams

Carefully assess the condition of fabrics and foams in the seating product, and have these replaced partially or even completely if
need be. In severe cases if the chassis is worn and the seating in poor condition it may be necessary to have the product re-furbished
by the manufacturer. In very severe cases if the chassis is badly worn and the seating in very poor condition then the product should
be decommissioned and disposed of.

Electrical and Electronic Equipment Decomissioning

If your product is set up with an electrical function you should always contact Sunrise Medical or your authorised representative for
de-comissioning information.

Appropriateness of Equipment

Check that the equipment supplied is appropriate for the needs of the user taking age, weight, ability, diagnosis, and any other
important factors into account. For instance, Sunrise Medical Seating Solutions Ltd. products may not be suitable for persons with
severe challenging patterns of behaviour. Also ensure safe specified limits for use are not exceeded by the new user. Your sales
representative can advise on the suitability of the equipment.

Manuals and Records

Ensure this instruction manual and any additional manuals for the seating unit are handed over to the new user with the equipment
and ensure the servicing log history is up to date. Records on the new user must be updated to maintain traceability.

Inspection

The equipment should have a rigorous inspection prior to the reissue, to include the points covered in the six monthly routine check,
regardless of if the checks were recently completed. The equipment should be free of excessive wear or distortion in any aspect of its
main construction or accessories.

Condition

Wheels, tyres and brakes must be in good working condition, properly adjusted and inflated as appropriate.
Neptune Rev.2.0