Herunterladen Inhalt Inhalt Diese Seite drucken

Linearization - KERN KFS-TM Betriebs Und Installationsanleitung

Vorschau ausblenden Andere Handbücher für KFS-TM:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6.6 Linearization

Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range. If linearity deviation is discovered during a monitoring of test resources, you
can improve this by means of linearization.
• Carrying out linearization is restricted to specialist staff possessing
well acquainted with the workings of weighing scales.
• The test weights to be used must be adapted to the weighing scale's
specifications; see chapter "monitoring of test resources".
• Observe stable environmental conditions. Stabilisation requires a
certain warm-up time.
• Do not remove the adjustment weight during linearization in step
LOAD 1 to LOAD 4, merely increase it instead. Conversely do not
remove the adjustment weight during step LOAD 4 to LOAD 1,
merely increase it instead.
• After successful linearisation you will have to carry out calibration;
see chapter "testing instruments control".
Tab. 1: Adjustment weights „LOAD1 – LOAD4"
MAX
LOAD 1
3kg
0.5kg
6kg
1kg
15kg
3kg
30kg
10kg
60 kg
10kg
120 kg
30kg
300 kg
50kg
600 kg
100kg
1.5 t
500kg
3 t
1000kg
KFS-TM-BA_IA-e-1916
LOAD 2
LOAD 3
1kg
2kg
2kg
4kg
5kg
10kg
15kg
20kg
20kg
40kg
60kg
60kg
100kg
200kg
200kg
400kg
750kg
1000kg
1500kg
2000kg
LOAD 4
3kg
6kg
15kg
30kg
60kg
150kg
300kg
600kg
1500kg
3000kg
17

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis