Herunterladen Inhalt Inhalt Diese Seite drucken

Measuring Capacity; Measuring Temperature Pt 100 / Pt 1000; Quick Peak Mode; Db Mode - Metrix Mtx3250 Bedienungsanleitung

50.000 punkte-tischmultimeter
Vorschau ausblenden Andere Handbücher für mtx3250:
Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 56
Technical specifications
Measuring
capacity
( )
Measuring
temperature
Pt 100 / Pt 1000
Quick PEAK
mode
@
Notes

dB mode

Resistive power

52
(cont.)
On this position, users can measure capacity of capacitor.
Rate
Precision
1% + 8D
50 nF (∗)
1% + 5D
500 nF (∗)
5 µF
1% + 3D
50 µF
1% + 3D
500 µF
1% + 3D
5000 µF
2% + 3D
50 mF
2% + 3D
(∗) A shielded cable is recommended for measurements below 50 nF.
Automatic or manual range selection.
5,000-count resolution.
"Active" Protection with CTP thermistance.
Open circuit maximum voltage: 9 V.
On this position, users can measure temperature via a Pt 100 / Pt 1000
sensor.
Rate
-125°C to + 75°C
-200°C to + 800°C
"Active" Protection with CTP thermistance.
Display in °C / K / °F available
Secondary magnitudes
Peak V
t > 500 µs
Peak A
t > 500 µs
• Resolution > 0.5 % [> 1 % 600 Vrms et 10 Arms range]
Results < 0.2 % are not displayed (1000 V < 100 D range).
Peak factor value is obtained by calculating: CF = VPk / Vrms
Displaying dBm measurement in relation to resistance reference that can be
adjusted from 1 Ω to 10,000 Ω, factory set to 600 Ω.
Resolution
dB absolute error
Additional calculation error
Measurement range
Protection
Displaying relative power measurement in relation to resistance reference
that can be adjusted from 1 Ω to 10.000 Ω, factory set to 600 Ω.
Range
Resolution
Precision
Protection
Measureme
Resolution
nt current
1 µA
1 µA
10 µA
100 µA
1 mA
1 mA
1 mA
Measurement
Resolution
current
0.1°C ≡ 2 µV
1 mA
0.1°C ≡ 2 µV
1 mA
Pk ranges
750 mV to 850 V
1000 V
750 µA to 30 A
0.01 dB
0.09 x V
0.01 dB
10 mV to 600 V
600 V
AC
DC and AC
100 µW
2 x accuracy in V
600 V
AC
Technical Specifications
Measurem
ent time
< 1.5 s
10 pF
< 1.5 s
100 pF
< 1.5 s
1 nF
< 1.5 s
10 nF
< 2 s
100 nF
≈ 3 s/mF
1 µF
≈ 3 s/mF
10 µF
Accuracy
±0.5°C
±1°C
Additional error
Protection
3% + 10 D
600 Veff.
3% + 18 D
10 Arms
4% + 10 D
relative error expressed in %
AC
/ V
(in %)
DC
AC
50,000-count Benchtop Multimeter
Protection
600 Veff.
Protection
600 Veff.
or
Inhaltsverzeichnis
loading

Inhaltsverzeichnis