5.7. Clear all values
Memory full
When the memory has stored 60 results, the display shows «Full M» after a
measurement. From this point onwards, a new measured value is stored by over-
writing the oldest value.
Clear all values
If you are sure that you want to permanently remove all stored values, hold down the
«Memory»-button (the instrument must have been switched off beforehand) until «CL» appears
and then release the button. To permanently clear the memory, press the «Memory»-button
while «CL» is flashing. Individual values cannot be cleared.
6. Error messages/malfunctions
If an error occurs during a measurement, the measurement is discontinued
Erreur n° Cause(s) possible(s)
Err 1
The systolic pressure was determined but afterwards the cuff pressure fell below
20 mmHg. This situation can occur, for example, if the tube becomes unplugged after
been detected.
Err 2
moved during the measurement (Artefact).
Err 3
Inflation of the cuff takes too long. The cuff is not correctly seated or the hose
connection is not sealed.
ERR 5
The measured reading indicated an unacceptable difference between systolic and
diastolic pressures. Take another measurement following directions carefully. Contact
your doctor if you continue to get unusual readings
39