Check whether the danger has passed or the problem has been resolved before
resetting the emergency stop.
The orange LED light on the hand control will flash when the emergency stop
has been activated.
4.2. Sleep setting
After around an hour, the chair will fall into a 'sleep' setting and the battery indication
LEDs on the handcontrol will go out. In order to use the chair again, it must be activated
(See paragraph 2.2 Activating).
4.3. Emergency lowering
If the controls stop working and the chair is in a raised posi-
tion, it can be lowered using the emergency lowering button.
Check first to make sure the chair is not in sleep setting!
• Remove the plug from the handcontrol (1) (See figure 18).
• Press the emergency lowering button (2) (See figure 18).
The chair will now return to the lowest position
(make sure no persons or objects are under the chair).
Take the client from the chair and make sure the device is not used until it has been
repaired. Place the chair in a safe place and indicate that it is out of order.
Contact the technical service or Lopital's service department. E-mail: service@lopital.nl
4.4. On / Off switch
On the right hand side of the cover cap, there is the on /
off switch (See figure 19). If the chair is out of use for an
extended period, you are advised to turn it off. This will
prevent any unauthorised use of the chair.