Herunterladen Inhalt Inhalt Diese Seite drucken

Aeroneb Solo Bedienungsanleitung Seite 198

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

note 1: At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
note 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
ª Field strengths from fixed transmitters, such as base stations
for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured
field strength in the location in which the Aeroneb Solo nebulizer
system is used exceeds the applicable RF compliance level above,
the Aeroneb Solo nebulizer system should be observed to verify
normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orientating or relocating
the Aeroneb Solo nebulizer system.
Over the frequency range 150 kHz to 80 MHz, field strengths
b
should be less than [3]V/m
Aeroneb
Solo System Instruction Manual
®

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Aeroneb Solo

Inhaltsverzeichnis