immunity test iec/en 60601
test level
Voltage
<5 % Ut
dips, short
(>95 % dip
interruptions
in Ut) for
and voltage
0.5 cycle
variations on
power supply
40 % Ut
input lines
(60 % dip
in Ut) for
IEC 61000-4-11
5 cycles
70 % Ut
(30 % dip
in Ut) for
25 cycles
<5 % Ut
(>95 % dip in
Ut) for 5 sec
Power
frequency
(50/60 Hz)
Magnetic field
3 A/m
IEC 61000-4-8
Note: Ut is the A.C. mains voltage prior to application of the
test level.
compliance
electromagnetic
level
environment –
Mains power
<5 % Ut
quality should be
(>95 % dip
that of a typical
in Ut) for
commercial
0.5 cycle
or hospital
environment. If
40 % Ut
the user of the
(60 % dip
Aeroneb Solo
in Ut) for
nebulizer system
5 cycles
requires continued
operation during
70 % Ut
power mains
(30 % dip
interruption, it is
in Ut) for
recommended that
25 cycles
the Aeroneb Solo
nebulizer system
<5 % Ut
be powered from
(>95 % dip in
an uninterruptible
Ut) for 5 sec
power supply or
battery.
Power frequency
magnetic fields
should be at levels
characteristic of a
3 A/m
typical location in a
typical commercial
or hospital
environment.
Aeroneb
Solo System Instruction Manual
®
guidance