Guidance and manufacturer's declaration – electromagnetic
The Aeroneb Solo nebulizer system is intended for use in the
electromagnetic environment specified below. The customer or the
user of the Aeroneb Solo nebulizer system should assure that it is
used in such an environment.
immunity test iec/en 60601
Electrostatic
±6 kV
discharge
contact
(ESD)
IEC 61000-4-2
±8 kV air
Electrical fast
±2 kV for
Transient/burst
power supply
lines
IEC 61000-4-4
±1 kV for
input/output
lines
Surge
±1 kV
differential
mode
IEC 61000-4-5
±2 kV
common
mode
Aeroneb
Solo System Instruction Manual
®
immunity
compliance
test level
level
±6 kV
contact
±8 kV air
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±1 kV
differential
mode
±2 kV
common
mode
electromagnetic
environment –
guidance
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Mains power
quality should be
that of a typical
commercial
or hospital
environment.