2. Safety
WARNING!
Observe the working conditions in accordance with chapter
■
3 "Specifications".
Prior to opening the instrument, disconnect it from the power
■
supply. Take care that, in fitting the instrument and connec-
tions, all components are protected against direct contact.
Observe standard regulations and safety rules for electrical,
■
low-power and high-power systems, especially any country-
specific safety regulations (for example VDE 0100).
Design the wiring particularly carefully when connecting
■
to other devices (e.g. a PC). Under certain circumstances,
internal connections in third-party devices (e.g. GND
connected to the safety earth) can lead to impermissible
voltages.
If the connected cable is longer than 30 metres, or it runs
■
outside the building, an additional and suitable overvoltage
protection should be provided.
The safety of the operator may be endangered if, for example:
there is visible damage to the instrument.
■
the instrument is not working as specified.
■
the instrument has been stored under unsuitable conditions for an
■
extended period of time.
If there is any doubt, please return the instrument to the manufacturer for
repair or maintenance.
WIKA operating instructions attachable indicator model A-IAI-1
EN
9