A. Instructions for use
1. InnoSpire Go needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in this document;
2. Portable and mobile RF communications equipment can affect InnoSpire Go.
B. Technical description
1. Warning that the use of accessories, transducers and cables other than those specified with the exception
of transducers and cables sold by the manufacturer of the InnoSpire Go as replacement parts for internal
components, may result in increased EMISSIONS or decreased IMMUNITY of the InnoSpire Go;
2. Warning that the InnoSpire Go should not be used adjacent to or stacked with other equipment.
3.
Guidance and manufacturer's declaration – electromagnetic emissions
The InnoSpire Go is intended for use in the electromagnetic environment specified below. The customer or
the user of the InnoSpire Go should assure that it is used in such an environment.
Emissions
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
flicker emissions
IEC61000-3-3
196
Electromagnetic environment –
guidance
The InnoSpire Go uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause any interference in nearby
electronic equipment.
The InnoSpire Go is suitable for
use in all establishments, including
domestic establishments and those
directly connected to the public
low-voltage power supply network
that supplies buildings used for
domestic purposes.
4.
Guidance and manufacturer's declaration – electromagnetic immunity
The InnoSpire Go is intended for use in the electromagnetic environment specified below. The customer or
the user of the InnoSpire Go should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electrostatic
±8 kV contact
±8 kV contact
discharge (ESD)
±15 kV air
±15 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
±2 kV for power
transient/burst
supply lines and
supply lines and
dc lines
dc lines
IEC 61000-4-4
Surge
±1 kV line(s)
±1 kV line(s) and
and neutral
neutral
IEC 61000-4-5
Voltage dips, short
<0 % U
<0 % U
T
interruptions and
(100 % dip in U
)
(100 % dip in U
T
for 0.5 cycle
for 0.5 cycle
voltage variations
on power supply
0 % U
0 % U
T
input lines
(100 % dip in U
)
(100 % dip in U
T
IEC61000-4-11
for 1 cycle
for 1 cycle
70 % U
70 % U
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles (50Hz)
for 25 cycles (50Hz)
for 30 cycles (60Hz)
for 30 cycles (60Hz)
0 % U
0 % U
T
(100% dip in U
)
(100% dip in U
T
for 250 cycles
for 250 cycles
(50Hz)
(50Hz)
for 300 cycles
for 300 cycles
(60Hz)
(60Hz)
Power frequency
3 A/m
N/A
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
ENGLISH
Electromagnetic environment – guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
The InnoSpire Go is suitable for use in
all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
The InnoSpire Go is suitable for use in
all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
The InnoSpire Go is suitable for use in
T
)
all establishments, including domestic
T
establishments and those directly
connected to the public low-voltage
T
power supply network that supplies
)
T
buildings used for domestic purposes.
T
)
T
T
)
T
N/A
197