Cleaning
After each use:
• Pour away residual medication from the medication chamber.
• Press the mouthpiece assembly release button to separate the mouthpiece assembly from
the handset.
• Rinse the mouthpiece assembly thoroughly under running tap water.
• Shake off excess water and allow to air dry fully before storing.
Daily cleaning
Wash the mouthpiece assembly by hand in a bowl of warm soapy water
(liquid dishwashing soap).
Rinse the mouthpiece assembly thoroughly under running tap water.
Shake off excess water and allow to air dry fully before storing.
Weekly disinfection
CAUTION: Risk of scalding. Use care around boiling water and in handling hot parts.
Boil the mouthpiece assembly in
water for 10 minutes. Ensure the
medication chamber lid is open
and there is enough water in the
pan to prevent the mouthpiece
assembly from touching the
bottom or boiling dry. Shake off
excess water and allow to air dry
fully before storing.
Or
• Disinfect the mouthpiece assembly in a Philips Avent baby bottle electric steam sterilizer, or
equivalent, for one cycle (follow the instructions of the steam sterilizer). Allow to air dry fully
before storing.
Or
• Immerse the mouthpiece assembly in a disinfectant of the Gluteraldehyde group
(testing performed with Korsolex Extra, 4% for 15 minutes).
• Wipe your handset clean with a clean damp cloth weekly or as needed. Do not use other
cleaning methods or solutions.
8
Mask cleaning
If using a mask, once a week, agitate in warm soapy water for 2 minutes or soak
for 10 minutes, rinse and air dry.
Maintenance
Mouthpiece assembly can be cleaned (boiled/disinfected) up to 52 times over 12 months.
To keep your InnoSpire Go working at the optimum level you should replace the mouthpiece
assembly and mask, if used, every 12 months as this is a consumable part.
LED Indication
LED Indication
SOLID GREEN
SOLID AMBER
FLASHES AMBER FIVE TIMES
FLASHES GREEN OR AMBER FIVE TIMES The treatment is finished
PULSES GREEN
ENGLISH
What it means
The battery is charged
There is enough charge for at least one
more treatment
There is not enough charge to take a treatment
The battery is being charged
9