P3 Pro
tri
CoUnt
rESEt
SEtGrA
* default setting

10 Error messages

Display
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
MBC-BAPS-e-1811
Not documented
Not documented
Reset weighing scale to factory setting
Not documented
Description
Zero range exceeded
(on start-up or when pressing the
• Load on weighing pan
• Excess load, during zero setting of weighing scale
• Incorrect adjusting process
• Fault on load cell
Value outside the A/D changer range
• Damaged weighing cell
• Damaged electronics
Lack of possibility of zero point initiation
• Damaged / overloaded measuring cell
• The objects are located on the platform / have contact
with it
• Unremoved transportation protection
• Main Board damaged
key)